IntroductionSelf-monitoring home blood pressure (BP) devices are currently recommended for long-term follow-up of hypertension and its management. Some of these devices are integrated with algorithms aimed at detecting atrial fibrillation (AF), which is common essential hypertension. This study was designed to compare the diagnostic accuracy of two widely diffused home BP monitoring devices in detecting AF in an unselected population of outpatients referred to a hypertension clinic because of high BP.MethodsIn 503 consecutive patients the authors simultaneously compared the accuracy of the Microlife® BP A200 Plus (Microlife) and the OMRON® M6 (OMRON) home BP devices, in detecting AF.ResultsSystolic and diastolic BP as well as heart rate (HR) values detected by the two devices were not significantly different. Pulse irregularity was detected in 124 and 112 patients with the OMRON M6 and Microlife BP A200 Plus devices, respectively. Simultaneous electrocardiogram (ECG) recording revealed that pulse irregularity was due to AF in 101 patients. Pulse irregularity detected by the OMRON M6 device corresponded to AF in 101, to supraventricular premature beats in 18, and to frequent premature ventricular beat in five patients, respectively. Pulse irregularity detected by the Microlife BP A200 Plus device corresponded to AF in 93, to supraventricular premature beats in 14, and to ventricular premature beats in five patients. The sensitivity for detecting AF was 100%, the specificity was 92%, and diagnostic accuracy 95% for the OMRON M6 and 100%, 92%, and 95 for the Microlife BP A200 Plus, respectively. AF was newly diagnosed by ECG recordings in 47 patients, and was detected in all patients by the OMRON device, and in 42 patients by the Microlife device.ConclusionThese results indicate that OMRON M6 is more accurate than Microlife BP A200 Plus in detecting AF in patients with essential hypertension. Widespread use of these devices in hypertensive patients could be of clinical benefit for the early diagnosis and treatment of this arrhythmia and related consequences.

Comparison of Microlife BP A200 Plus and Omron M6 Blood Pressure Monitors to Detect Atrial Fibrillation in Hypertensive Patients

Maurizio M Volterrani;Lombardo M;
2012-01-01

Abstract

IntroductionSelf-monitoring home blood pressure (BP) devices are currently recommended for long-term follow-up of hypertension and its management. Some of these devices are integrated with algorithms aimed at detecting atrial fibrillation (AF), which is common essential hypertension. This study was designed to compare the diagnostic accuracy of two widely diffused home BP monitoring devices in detecting AF in an unselected population of outpatients referred to a hypertension clinic because of high BP.MethodsIn 503 consecutive patients the authors simultaneously compared the accuracy of the Microlife® BP A200 Plus (Microlife) and the OMRON® M6 (OMRON) home BP devices, in detecting AF.ResultsSystolic and diastolic BP as well as heart rate (HR) values detected by the two devices were not significantly different. Pulse irregularity was detected in 124 and 112 patients with the OMRON M6 and Microlife BP A200 Plus devices, respectively. Simultaneous electrocardiogram (ECG) recording revealed that pulse irregularity was due to AF in 101 patients. Pulse irregularity detected by the OMRON M6 device corresponded to AF in 101, to supraventricular premature beats in 18, and to frequent premature ventricular beat in five patients, respectively. Pulse irregularity detected by the Microlife BP A200 Plus device corresponded to AF in 93, to supraventricular premature beats in 14, and to ventricular premature beats in five patients. The sensitivity for detecting AF was 100%, the specificity was 92%, and diagnostic accuracy 95% for the OMRON M6 and 100%, 92%, and 95 for the Microlife BP A200 Plus, respectively. AF was newly diagnosed by ECG recordings in 47 patients, and was detected in all patients by the OMRON device, and in 42 patients by the Microlife device.ConclusionThese results indicate that OMRON M6 is more accurate than Microlife BP A200 Plus in detecting AF in patients with essential hypertension. Widespread use of these devices in hypertensive patients could be of clinical benefit for the early diagnosis and treatment of this arrhythmia and related consequences.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.12078/2759
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 59
  • ???jsp.display-item.citation.isi??? 57
social impact